Epson C82302/04 (Serial I/F) Bedienungsanleitung Seite 76

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 294
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 75
2-20
3
Press the
OPTION
key twice to move the cursor under DIAG and
press
ENTER
.
A display referring to the first diagnostic test appears:
4
Change the "1" to a "2" by pressing
2.
Then press
ENTER.
The following display appears and the test begins:
If Heater Test Passed appears on the display as shown below,
proceed to the chiller test. If Heater Test Failed appears on the
display, notify an Applied Biosystems Service Engineer.
5
Press
STOP
to go to the next test (Test #3).
Next, you will run the chiller test, which is accessible from the first
diagnostic display.
Enter Diag test # 1
REVIEW HISTORY FILE
Enter Diag test # 2
HEATER TEST
Heater Test Passed
Seitenansicht 75
1 2 ... 71 72 73 74 75 76 77 78 79 80 81 ... 293 294

Kommentare zu diesen Handbüchern

Keine Kommentare