8719/20/22ET Specifications and Characteristics
Instrument Specifications
Table2-62.4-mm (50) Device Connector Type
8722ET (2.4-mm, 50)
Description
Specification
50 to 500 MHz
0.5 to 2 GHz 2 to 8 GHz 8 to 20 GHz 20 to 40 GHz
Directivity
a
(dB)
23 23 21 16 15
Source Match
a
(dB)
16 20 14 11 10
Load Match
a
(dB)
22 20 20 20 15
Reflection
b
Tracking (dB)
±3 ±3 ±3 ±3 ±4
Transmission
a
Tracking (dB)
±3 ±3 ±3 ±3 ±3
Tracking
Stability (Ratio
Measurement)
Magnitude
(dB)
0.02/˚C, typ. 0.03/˚C, typ. 0.03/˚C, typ. 0.04/˚C, typ. 0.06/˚C, typ.
Phase (deg) 0.1/˚C, typ. 0.1/˚C, typ. 0.3/˚C, typ. 0.5/˚C, typ. 1.0/˚C, typ.
Crosstalk
c
(dB)
95 95 95 88 82
a.Does not include effects of cable on test ports.
b.Doesnotincluderolloffbelow500MHzwhichistypically18dBat100MHz,and25dBat
50 MHz.
c.Measurement conditions: Normalized to a thru, measured with two shorts, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of the
maximum power out or the maximum receiver power.