Seitenansicht 115
ZVx SENSe Subsystem
1043.0009.50 3.77 E-15
[:SENSe[1 to 4]:]AVERage:MODE
The command selects the type of the average function.
Syntax: [:SENSe[1 to 4]:]AVERage:MODE SWEep | POINt
Example: "AVER:MODE POIN"
Features: *RST value: NORMal
SCPI: device-specific
3.6.14.2 SENSe:BANDwidth Subsystem
This subsystem controls the setting of the analyzers filter bandwidths. The commands BANDwidth and
BWIDth have the same effect.
COMMAND PARAMETER UNIT COMMENT
[SENSe<1...4>]
:BANDwidth
[:RESolution]
:BWIDth
[:RESolution]
<numeric_value>
<numeric_value>
HZ
HZ
[SENSe[1 to 4]:]BANDwidth|BWIDth[:RESolution]
This command defines the analyzers resolution bandwidth.
Syntax: [SENSe[1 to 4]:]BANDwidth|BWIDth[:RESolution] <numeric_value>
<numeric_value>::= 1kHz...26kHz; "MAX" must be entered instead of 26 kHz.
Example: "BAND 10kHz"
Features: *RST value:
SCPI: conforming
Seitenansicht 115